The brand new acoustic profiler joining the SeaGuard family. It features innovative development of the acoustic profiling capacity and an exceptional ability to collect high quality current information on moving and tilting moorings. Available as a self-recording instrument, it also integrates unique real time features to meet your application needs.
The SeaGuardII is a smart data hub that combines the SeaGuard electronics with the advanced management firmware of Aanderaa SmartGuard data hub.
SeaGuardII DCP is a 600kHz frequency profiler with multi-sensor capability. By design, we offer increased deployment time, optimized configuration flexibility and unique features to cope with demanding upper ocean environments.
Optional parameters are available using Aanderaa range of smart sensors that include temperature, pressure, conductivity, oxygen, wave, tide and turbidity. In addition the SeaGuardII has 4 analog inputs, 2 serial ports with power control and direct connection for real time data transmission unit.
Applications:
Increased deployment time:
Velocity profile measurement | ||
Acoustic frequency:
|
600 kHz
|
|
Typical profiling range:
|
Broadband: 30-70m
|
|
Narrowband 35-80m
|
1)
|
|
Cell size:
|
0.5m - 5m
|
|
Cell overlap:
|
0-90%
|
|
Velocity range:
|
Narrowband: 0-500 cm/s other range on request
|
|
Broadband:
|
0-400 cm/s | |
Velocity accuracy:
|
0.3 cm/s or ±1,5% of reading
|
|
Velocity resolution:
|
0.1 cm/s | |
Velocity precision:
|
<3,3cm |
2)
|
Ping rate:
|
Up to 10Hz | |
Cell positioning:
|
Static (instrument referred)
|
3)
|
Dynamic (surface referred)
|
||
Multiple columns
|
||
Number of columns:
|
3 simultaneous columns + Surface cell
|
3) |
Max. number of cells:
|
150 total, 75 for first column, 50 for second and 25 for third
|
|
Blanking zone:
|
1m
|
|
Transducers
|
||
Number of beams:
|
4
|
|
table:
|
Advanced autobeam algorithm
|
4)
|
Beam angle:
|
25°
|
|
Beam width:
|
2.5°Echo intensity
|
|
Echo Intensity
|
||
Dynamic range:
|
> 50dB
|
|
Resolution:
|
< 0.01dB
|
|
Precision:
|
< 0.01dB
|
|
Tilt and compass
|
||
Type:
|
Internal solid state
|
|
Pitch / roll range:
|
± 90° |
5)
|
Tilt / Heading accuracy:
|
± 1.5° / ± 3.5°
|
|
Tilt / Heading resolution:
|
< 0.1°
|
|
Embedded temperature sensor
|
||
Range
|
-4- +40°C
|
|
Resolution
|
0,001°C
|
|
Accuracy
|
± 0,05°C
|
|
Response Time (63%):
|
<5 sec
|
|
Communication and recording
|
||
Data storage:
|
2GB SD Card exchangeable and remote downloadable
|
|
Remote operation:
|
Device layout
|
|
Configuration
|
||
Recording start/stop
|
||
Status monitoring
|
||
Available telemetry
|
||
Cable, radio modem, GPRS,
|
||
GOES,
|
||
GSM, Iridium
|
||
Configuration and real time data software:
|
||
RT Collector (for Windows®XP, Windows®7)
|
||
Configuration interface:
|
USB / RS232 / RS422
|
|
Recording system:
|
Multiple sensors groups with
|
|
individual recording interval
|
||
and activation, immediate,
|
||
aligned or delayed start.
|
||
Recording interval:
|
From 30sec to 3hrs
|
|
Power
|
External power supply:
|
|
12-30V
|
||
Internal battery:
|
||
2 batteries inside the instrument:
|
Alkaline 3988: 9V, 15Ah
|
|
Lithium 3908: 7V, 35Ah
|
||
Current drain example:
|
4,2 mA
|
6)
|
Environmental
|
||
Depth rating:
|
300m
|
|
Operating temperature:
|
-5 to +40°C
|
|
Dimensions:
|
D: 160mm
|
|
H: 585mm
|
||
Weight:
|
||
In Air
|
10,8 kg
|
|
In Water
|
3,6 kg
|
|
Materials:
|
PET, PUR, Titanium, Stainless
|
|
steel 316
|
||
Optional sensors
|
||
Temperature Sensor 4060
|
||
Range:
|
-4-36°C (32-96.8°F)
|
7)
|
Resolution:
|
0.001°C (0.0018°F)
|
|
Accuracy:
|
±0.03°C (0.054°F)
|
|
Response Time 63%:
|
< 2 seconds
|
|
Conductivity Sensor 4319
|
||
Range:
|
0-7.5 S/m
|
|
Resolution:
|
0.0002 S/m
|
|
Accuracy
|
4319 A:
|
|
±0.005 S/m
|
||
4319 B:
|
||
±0.0018 S/m
|
||
Response Time:
|
<3s
|
8)
|
Pressure Sensor 4117
|
||
Resolution:
|
<0.0001% FSO
|
|
Accuracy:
|
±0.01% FSO
|
|
4117A Range:
|
0 – 1000kPa (0 – 145 psia)
|
|
4117B Range:
|
0 – 4000kPa (0 – 580 psia
|
|
Wave and Tide Sensor 5217/5218
|
||
Resolution :
|
<0,0001% FSO
|
|
Accuracy:
|
±0,01% FSO
|
|
Sampling rate:
|
2Hz, 4Hz
|
|
Tide:
|
5217 Range:
|
|
0 – 400kPa (0 – 58 psia)
|
||
5217A Range:
|
||
0 – 1000kPa (0 – 145 psia)
|
||
5217B Range:
|
||
0 – 4000kPa (0 – 580 psia
|
||
Wave:
|
No. of samples:256, 512, 1024, 2048
|
|
5218 Range:
|
||
0 – 400kPa (0 – 58 psia)
|
||
5218A Range:
|
||
0 – 1000kPa (0 – 145 psia)
|
||
Turbidity Sensor 4112:
|
0-5V Analog Output
|
|
4 models:
|
0-25, 0-125, 0-500, 0-2000FTU
|
|
Oxygen Optode 4835/4330
|
||
Concentration Air Saturation
|
||
Measurement Range:
|
0 – 500 μM 0 - 150%
|
|
Resolution:
|
< 1 μM 0.4%
|
|
Accuracy:
|
<8 μM or 5%
|
10)
|
With multipoint calibration:
|
<±2 μM or ±1.5% | |
Response Time (63%):
|
4330F (fast response foil) <8 sec
|
|
4835/4330 (standard foil) <25 sec
|
||
Analog and serial inputs:
|
||
Analog:
|
4 channels 0-5V | |
Serial:
|
2 channels with sensor and power
|
|
switching one RS232 port and one
|
||
RS422
|
12)
|
|
1. Typical range with normal backscatter conditions. The measurement range is highly dependent on the scattering conditions. For waters with low amount of scatters, expect a shorter range than for waters with a high amount of scatters
|
||
2. Standard deviation For the horizontal velocity in broadband mode, 3m cell size
|
||
3. Requires pressure sensor 4117
|
||
4. Optimal selection of the the best 3-beams conbination to avoid interference and beam failures
|
||
5. Compensation calibrated up to ± 35°
|
||
6. In Broadband mode, 30min interval, 20*2 pings, 2m cell size, 20 cells
|
||
7. Extended range available on request.
|
||
8. Dependent on flow through cell bore
|
||
9. Multipoint calibration available on request: 40 points at 5 temperatures and 8 oxygen concentrations
|
||
10. Requires salinity compensation for salinity < 1mS/cm
|
||
11. Within calibrated range 0-120%
|
||
12. The serial ports may be used either as serial sensor inputs or serial real-time output
|