The 1088 Rotary Autosampler is specifically designed to operate with the Aurora 1030W TOC analyzer or Aurora 1030C TOC analyzer. The 1088 autosampler aspirates liquid samples from vials and transfers each sample aliquot to an Aurora 1030 TOC analyzer for analysis. A removable 88-position autosampler tray loads the sample vials for fully automated, unattended operation.
The 1088 supports a number of special functions to address a range of sample conditions and analysis requirements. Onboard magnetic stirring ensures samples containing insoluble components or particles are homogeneous for sampling and accurate analysis. Septum piercing provides closed vial sampling for applications where open vials may comprise the accuracy of measurements. System configuration for sample pretreatment allows pre-acidification and purging of samples within the 1088 prior to sampling. This technique removes the TIC content of samples, reducing the analysis time required for TOC measurements.
Dimensions (1088 Autosampler) | 64.77–cm H x 38.1–cm W x 48.26–cm D (25.5" H x 15" W x 19" D) |
Weight (1088 Autosampler) | 19 kg (42 lbs) |
Dimensions (1088 Autosampler & Aurora 1030) |
66–cm H x 47–cm W x 66–cm D (26" H x 18.5" W x 26" D) |
Weight (1088 Autosampler & Aurora 1030) | 34.4 kg (76 lbs) |
Benchspace Footprint | 3,102cm2 (494 in2) |
Environment | 15–35 °C operating temperature 10–90% relative humidity |
Sample Capacity | 88 sample vials |
Sampling Needle | 8.4", two-hole, stainless steel |
Sample Transfer Tubing | 0.045" I.D. x 1/8" O.D. Teflon® tubing |
Vials | 40-mL VOA vials |
Caps | Open-hole caps |
Septa | TFE-faced septa |
Sample Transfer Volume Range | 10 μL–10 mL |
Particulate Handling | Up to 500 μm |
Special Functions | On-board magnetic stirring Septum piercing Sample pre-treatment for TIC removal |
Power Supply | 100–230 VAC (±10%); 50/60 Hz; 150 W (max) |
Communication | RS-485 (Aurora 1030 analyzer to 1088 Autosampler) |
Certifications | CE |
Warranty | 12 months on parts and labor |